Technical Description
3-10
Operating instructions ULTRAMAT 23 gas analyzer
C79000-B5276-C216-03
3.4 Integration
Communication
Communications facilities
The gas analyzers of series 6 (ULTRAMAT 6,
ULTRAMAT/OXYMAT 6, OXYMAT 6/61/64, FIDAMAT 6 and
CALOMAT 6/62), as well as the ULTRAMAT 23 offer the following
communications facilities:
RS 485 interface
PROFIBUS DP/PA
AK interface (only OXYMAT 6, ULTRAMAT 6 and
ULTRAMAT/OXYMAT 6).
RS 485 interface (ELAN)
The standard integrated series interface permits to communi-
cate between several gas analyzers via the internal bus (ELAN).
Up to 12 gas analyzers with max. four components can be net-
worked.
The operating principle is shown in the following figure.
Typical design of an RS 485 network
.Interface parameters
RS 485 interface (SIPROM GA)
SIPROM GA is a software tool for service and maintenance. All
analyzer functions (except factory functions) can be remote-
controlled and monitored via RS 485/RS 232 converters,
RS485/ethernet converters and RS 485/USB converters.
Up to 12 gas analyzers with max. four components can be net-
worked.
Networking of several gateways is possible when using the
RS 485/Ethernet converter (gateway). The number of operatable
analyzers is increased correspondingly.
Functions
Display and saving of all analyzer data
Remote operation of all analyzer functions
Parameter and configuration settings
Comprehensive diagnostics information
Remote calibration
Online help
Cyclic saving of measured values and status on hard disk
Exporting to commercially available application programs
Downloading of new software.
Hardware requirements:
PC/Laptop Pentium 133 MHz, RAM 32 MB, CD-ROM drive
Free disk capacity min. 10 MB
Free COM-Port: RS 232 or RS 485:
- for coupling to ELAN network RS 485/RS 232
- Linking the RS 485/Ethernet converter requires a 10 MB
standard network (RJ 45 connection) with TCP/IP. The dis-
tance should not exceed 500 m with an RS 485 network, a
repeater should be inserted for a longer distance.
Software requirements:
•Windows 95
•Windows 98
Windows NT 4.0
Windows 2000
Windows XP.
Item Designation
1Computer
2 RS 485/RS 232 converter with RS 485/RS 232 cable
3 RS 485 bus connector with jumper
4 Analyzers
5RS485 cable
6 RS 485 bus connector
7 RS 485 network
8 9-pin SUB-D plug
9 Option: RS 485 repeater
Level RS 485
Baud rate 9600
Data bits 8
Stop bit 1
Start bit 1
Parity None
No echo mode
Ordering information Order No.
Interface description A5E00054148
RS 485/RS 232 converter
C79451-Z1589-U1
RS 485/USB converter
A5E00852382
Further accessories (e.g. cable,
connectors, repeater, etc.)
see Catalog IK PI