Laboratory centrifuge Sigma 2-7, Sigma 2-7 Cyto
Laboratory centrifuge Sigma 2-7 IVD
6 Using the centrifuge
28 / 68
6.2.2.2 Installation of accessories
Only use inserts that are suitable for the rotor.
All buckets of the swing-out rotor need to be installed when spinning.
Always load the axial symmetrical inserts/buckets of the rotors with the
same accessories and fill to avoid imbalance.
Centrifugation with different tube sizes
Working with different tube sizes is possible. In this case, however, it is
very important that axial symmetrical inserts are identical (see figure).
correct
incorrect
Fig. 5: Permissible and impermissible loading of a swing-out rotor with different tube sizes
(example illustration)
Centrifugation with low capacity
Install the tubes axial symmetrically so that the buckets and their inserts
are loaded evenly.
It is not permissible to load angle rotors on only one axis.
correct
incorrect
Fig. 6: Permissible and impermissible loading of an angle rotor and a swing-out rotor
(example illustrations)