Laboratory centrifuge Sigma 2-7, Sigma 2-7 Cyto
Laboratory centrifuge Sigma 2-7 IVD
7 Malfunctions and error correction
38 / 68
7.1.1 Emergency lid release
In the event of a power failure, it is possible to manually open the centrifuge
lid.
Switch off the mains power switch and disconnect the power cord from
the socket.
Remove the plug (see figure, item 1) from the front panel, e.g. with a
screwdriver.
Fig. 10: Position of the opening for the emergency lid release
Insert the supplied hexagon socket key horizontally into the hole and
turn it anti-clockwise to the stop. The lid lock will then audibly unlock.
Fig. 11: Manual release of the lid lock
Then, reinsert the plug.
WARNING
Do not unlock or open the lid unless the rotor is at a standstill.
If the lid is opened via the emergency lid release system during a centrifuge
run, the centrifuge will be switched off immediately and decelerate in an
unbraked manner.