11.3 Table of the service life of rotors and accessories
• If no other data concerning the service life are engraved on the rotor or accessory, rotors and
buckets must be checked by the manufacturer after 10 years.
• If a specification concerning the maximum number of cycles and a specification concerning the
service life (i.e. a date) are provided, the specification that occurs first shall apply.
• After 50,000 cycles, rotors must be scrapped for safety reasons.
Service life
("Exp.Date")
Autoclaving Suitable for centrifuge Remarks
4-15C, 4K15C, 4-16, 4-16S,
4-16K, 4-16KS, 6-15, 6K15,
6-16, 6-16K
without engraving, only
"spincontrol professional"
and "spincontrol S"
*in combination with
bucket 13850 or 13860
1-16 Edition, 1-16K Edition
1-15, 1-15K, 1-15P, 1-15PK
1-15, 1-15K, 1-15P, 1-15PK
1-15, 1-15K, 1-15P, 1-15PK
4-16, 4-16S, 4-16K, 4-16KS,
6-16, 6-16S, 6-16K, 6-16KS
2-6, 2-6E, 2-7,
2-16P, 2-16KL, 2-16KHL
2-6, 2-6E, 2-7,
2-16P, 2-16KL, 2-16KHL
6-16, 6-16K, 6-16S, 6-16KS
4-5L, 4-16S, 4-16KS,
4-16KHS, 6-16S, 6-16HS,
6-16KS, 6-16KHS
*in combination with
rotor 11805
*in combination with
rotor 11805
8K, 8KB, 8KS, 8KBS without engraving
8K, 8KB, 8KS, 8KBS without engraving
8K, 8KB, 8KS, 8KBS without engraving
8K, 8KB, 8KS, 8KBS without engraving