16 Setting Up the Pump
SIGMA Manual Number 41018
Revision AR
View Information
Pump Information – The pump information screen has the following read-only information
that either the user (clinician) or Bio-Medical Engineer may find useful. From the Pump
information screen, press the sw info softkey to display the Software Version screen, which
show the versions of the individual software components that are installed.
SW Version
Sharp build
PIC build
CPLD build
SmartBatt Charger build
Network Module build
Serial number – The serial number assigned by SIGMA for tracking and device history.
Tube type – the name of the IV tube set manufacturer that the device is calibrated for
use with.
Wireless Module Type
Library Information The library information screen identifies the name of the Active Drug
Library in pump memory as well as the date modified, version number, and format indica-
tors.
Show Clinical Advisory – If the current infusion has a clinical advisory associated with it,
selecting this option will display the advisory. If the current infusion does not have a clinical
advisory associated with it, this option is not selectable and appears in grey text in the dis-
play.
Infusion Information – The infusion information screen identifies the infusion-specific pri-
mary and secondary bag parameters that are not otherwise displayed on the setup screen.
The information includes the following information for each bag (if programmed): audio
level, near empty alert status (primary only), pressure setting, KVO rate (primary only), and
primary siphoning alert or secondary complete alert status (secondary only).
History Log – Selecting this option provides access to the event log. Available options for
viewing the log are listed below.
NOTE: Access to this option is provided only when the pump is not running
(delivering). If the pump is running, this option is not selectable and
appears in grey text in the display.
View History Log – select this option to view the entire history log on the pump screen.
View System Error Log – select this option to view only the system errors recorded in
the history log.
View Drug Error Log – select this option to view only the events associated with pro-
gramming an infusion; including any drug limits that may have been exceeded during
setup.
Dump History Log – select this option to send the entire history log out of the pump via
the IrDA port (refer to the Logging section of the SIGMA Spectrum Service Manual).