54 Cleaning and Storage
SIGMA Manual Number 41018
Revision AR
CAUTION:
- Alcohols are flammable and should not be used for Standard Battery/Wireless
Battery Module cleaning/disinfection. Always use alcohols in a well-ventilated
area.
- When cleaning the Standard Battery/Wireless Battery Module, care should be
taken to prevent shorting of the pack’s exposed terminals.
- Do not sterilize this device by autoclaving or ETO gas.
- Do not immerse any part of this device or allow cleaning fluids to seep inside the
pump.
- Do not use phenol-based cleaners/disinfectants. Phenols degrade plastics and
membrane switches. Phenols are intended for cleaning of hard non-porous
surfaces such as: sinks, counter tops, and stainless steel.
- Do not use abrasive cleaners.
Storing the Pump
Connect the AC power adaptor to the pump and supply source power to charge the pump’s
battery during storage. This insures a fully charged battery for subsequent use.
Do not store or transport pumps in ways that might result in physical damage.
For extended periods of storage remove the battery and repackage the pump in the original
shipping container.
Storage at elevated temperatures will diminish battery life.
Refer to Specifications section of this manual for recommended storage temperature and
humidity.
Servicing the Pump
CAUTION: Electric shock hazard.
Refer servicing to SIGMA trained and qualified service personnel. Refer to the
Service Manual for inspection and maintenance procedures.