54 Cleaning and Storage
SIGMA Manual Number 41018
Revision AR
CAUTION:
- Alcohols are flammable and should not be used for Standard Battery/Wireless
Battery Module cleaning/disinfection. Always use alcohols in a well-ventilated
area.
- When cleaning the Standard Battery/Wireless Battery Module, care should be
taken to prevent shorting of the pack’s exposed terminals.
- Do not sterilize this device by autoclaving or ETO gas.
- Do not immerse any part of this device or allow cleaning fluids to seep inside the
pump.
- Do not use phenol-based cleaners/disinfectants. Phenols degrade plastics and
membrane switches. Phenols are intended for cleaning of hard non-porous
surfaces such as: sinks, counter tops, and stainless steel.
- Do not use abrasive cleaners.
Storing the Pump
■ Connect the AC power adaptor to the pump and supply source power to charge the pump’s
battery during storage. This insures a fully charged battery for subsequent use.
■ Do not store or transport pumps in ways that might result in physical damage.
■ For extended periods of storage remove the battery and repackage the pump in the original
shipping container.
■ Storage at elevated temperatures will diminish battery life.
■ Refer to Specifications section of this manual for recommended storage temperature and
humidity.
Servicing the Pump
CAUTION: Electric shock hazard.
Refer servicing to SIGMA trained and qualified service personnel. Refer to the
Service Manual for inspection and maintenance procedures.