SIGMA Manual Number 41018
Revision AR
APPENDIX E - ELECTROMAGNETIC
COMPATIBILITY
Emissions
WARNING: The use of accessories or cables other than those specified by SIGMA
may result in increased Emissions or decreased Immunity of this medical
device.
Guidance and manufacturer’s declaration – electromagnetic emissions
The SIGMA Model Spectrum Infusion pump is intended for use in the electromagnetic environment speci-
fied below. The customer or user of the Spectrum should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1 The Spectrum uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B The Spectrum is suitable for use in all establish-
ments, including domestic establishments and
those directly connected to the public low-voltage
power supply network that supplies buildings used
for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/ flicker
emissions
IEC 61000-3-3
Complies