40 Programming the Pump
SIGMA Manual Number 41018
Revision AR
Software version 6.01.00 and later:
To titrate flow rates and VTBI without stopping the pump:
1. Press the TITRATE softkey.
2. To titrate dose, press DOSE softkey.
3. To titrate rate mL/hr, press mL/hr softkey.
4. Observing the displayed hard and soft dose or mL/hr flow rate limits, enter a new dose or
mL/hr flow rate.
5. Press OK.
If soft limits are exceeded, press YES to accept new value or NO to return to prior value.
6. To titrate VTBI, press VTBI softkey, enter new value, and press OK.
ICU
10
optionsreview
DOPamine*
mcg/kg/min
titrate
bolus