48 Alarms
SIGMA Manual Number 41018
Revision AR
Downstream Occlusion Eliminate a closed clamp, kinked tube, positional catheter,
clotted catheter, clogged IV filter or other sources of occlusion
below the pump.
Downstream Pressure Limit –
Reset Setting
The downstream pressure limit differs from care area default
(set in Master Drug Library). Select YES to reset or NO to keep
the current setting.
Inactivity Alarm The pump has been inactive for 2 minutes and no action has
been taken. Follow the prompted action and resume or restart
the pump by pressing RUN.
Infusion Complete The VTBI (volume to be infused) has counted down to zero and
has been delivered. The pump is running at a rate of 1.0mL/hr
(KVO rate) – keep vein open (or the actual infusion rate,
whichever is lower). Press STOP to halt the KVO rate and
return to the setup screen. Select a new VTBI value and press
RUN.
In Stop – Slide Clamp Closed Open slide clamp and push RUN (or unload set).
Line Flush Complete The line flush VTBI has counted down to zero mL. the pupmp is
running at a KVO rate of 1 mL/hr. Press STOP to halt the KVO
rate and return to flush prompt. (Maximum 3 line flushes are
allowed).
This alarm applies to software versions 6.03.01 and later
infusing a primary or secondary drug in amount/time modes
only.
Low Battery The low battery alarm threshold has been reached. Plug the AC
Power Adaptor into the pump and into the AC source outlet as
soon as possible to recharge the battery.
Slide Clamp Closed Open slide clamp and press RUN or reload the set.
System Error An internal fault has been detected. Some faults can be cleared
by either cycling power (off, then on) or by turning the power
off, disconnecting the battery, reconnecting it several seconds
later and pressing the ON key. If neither procedure clears the
fault, return the pump for service.
Upstream Occlusion Eliminate the occlusion by checking for an upstream closed
clamp, kinked tube or closed burette valve and press the RUN
key.