66 APPENDIX A - Specifications
SIGMA Manual Number 41018
Revision AR
IrDA® Serial Infrared Physical Layer Link Specification v1.4(IrPHY), IrDA Serial Infrared
Link Access Protocol v1.1 (IrLAP) and IrDA Serial Infrared Link Management Protocol v1.1
(IrLMP), IrDA Tiny TP v1.1
Wireless – 802.11b, 802.11g
EIA-RS-232 levels for Asynchronous Transmit/Receive only (RS232)
Specification Description
Overall Size (Pump) With Standard Battery
Without IV pole clamp – 5.8” H x 4.2” W x 2.5” D
With IV pole clamp – 5.8” H x 6.4” W x 4.7” D
With Wireless Battery Module
Without IV pole clamp – 6.3” H x 4.2” W x 2.5” D
With IV pole clamp – 6.3” H x 6.4” W x 4.7” D
Weight With Standard Battery
Without IV pole clamp – 25.5 oz ± 1.0 oz
With IV pole clamp – 33.5 oz ± 1.0 oz
With Wireless Battery Module
Without IV pole clamp – 26.5 oz ± 1.0 oz
With IV pole clamp – 34.5 oz ± 1.0 oz
Volumetric Accuracy Accuracy is based on volume collected over one hour using
compatible Baxter and Hospira Standard IV Sets.
Specified accuracy is maintained on Baxter Standard IV Sets
for up to 96 hours (maximum 12 liters) and on Hospira
Standard IV Sets for up to 72 hours (maximum 9 liters)
Infusion Modes Continuous primary and secondary, Multi-Step, Cyclic TPN,
Amount/Time
BAXTER HOSPIRA
0.5 – 1.9 mL/hr ±0.1mL/hr ±0.1mL/hr
2.0 – 800 mL/hr ±5% ±5%
801 – 999 mL/hr ±5% ±10%