70 APPENDIX A - Specifications
SIGMA Manual Number 41018
Revision AR
Alarms and Alerts
Air-In-Line: dual beam ultrasonic detector alarms for
large bubbles but allows smaller bubbles to pass. Detects
air bubbles > 1” (» 125µL Hospira, » 140µL Baxter), will
alarm if > 1 mL* of air in 15 min.,
< 50µL bubbles are omitted in the summation of the 1
mL.*
*up to 1.5mL at 60°F
Audio: speaker actuated audio alarm, low, medium
and high levels selected through the User Settings, Alarm
Settings screen
Battery Missing – pump does not detect battery
attached
Depleted Battery: pump stops running, alarms for 3
minutes
Dose Rate Limit Exceeded: the pump will run with a
soft dose rate exceeded after a double confirmation, it will
not run with a hard dose rate exceeded, rates must be reset
within hard limits
Downstream Occlusion: automatic restart occurs after
the downstream occlusion is cleared. Actuation can be set
to Low, 6 ± 4 PSI, Medium, 13 ± 6 PSI or High, 19 ±9 PSI
Inactivity: actuates after the pump has been inactive
for 2 minutes
Infusion Complete: occurs when the VTBI reaches
zero, at which time a KVO rate begins (The previous
running rate or 1 mL/hr, whichever is lower)
In Stop – Load Set
In Stop – Open Slide Clamp
In Stop – Push Run
Low Battery - < 30 minutes of battery power remain
Shut Door
Slide Clamp Closed – Pump Stopped
System Error
Upstream Occlusion
Very Low Battery - <15 minutes of battery power
remain
Due for inspection: Preventative Maintenance and/or
Network Certification
Alarm Volume Variable (three levels: high, medium and low)