EasyManua.ls Logo

Sigma sigmaspectrum 35700ABB - Hardware Labeling

Sigma sigmaspectrum 35700ABB
102 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3
SIGMA Manual Number 41018
Revision AR
WARNING: Prevent Inaccuracy
The following can cause flow rate inaccuracies and must be avoided:
- Incompatible brand IV sets and compatible brand IV sets with
unusually large or small diameters or unusually stiff materials.
- Operating temperatures outside of 60-90°F for Standard Battery and
60-80°F for Wireless Battery Module.
- Using IV sets longer than is recommended in the Specifications section
of this manual.
- Using a dropped, damaged, dirty or wet pump.
- Pressurizing IV bags.
- Positioning IV containers more than 3 feet above or 1 foot below the
pump.
- Using Microdrip or Minidrip chambers for flow rate settings greater
than 200 mL/hr. Doing so may influence flow rate accuracy and cause
nuisance air or upstream occlusion alarms.
- Exceeding 500 mL/hr flow rate settings when using sets with
backcheck valves. Doing so may influence flow rate accuracy or cause
nuisance upstream air or upstream occlusion alarms. Flow rates above
300mL/hr may cause fluid to be siphoned from the primary container
during piggyback operation (see Secondary Infusion).
Note: Upstream occlusion detection is only effective for occlusions
present immediately after the start of the pump’s run operation.
Upstream occlusions caused by non-vented IV sets used with non-vented
glass bottles or closed burette air vents cannot be detected because of the
very slow-building vacuums resulting from these situations.
WARNING: This equipment is not suitable for use in the presence of a Flammable
Anesthetic Mixture with Air or with Oxygen or Nitrous Oxide. (This
statement is a requirement of the IEC—60601-2-24 standard. It applies
to oxygen enriched environments, such as oxygen tents. It is not meant to
apply to patients on breathing tubes.)
WARNING: Follow Epidural Precautions
Epidural administration of drugs other than those indicated for epidural
use can result in serious patient injury.
- When administering epidural analgesics, use only catheters specifically
labeled for epidural analgesia drug delivery.
- To help prevent accidental infusion of non-epidural drugs, DO NOT
USE epidural administration sets that contain injection sites.
- Label the administration container and IV set “EPIDURAL USE
ONLY”.
- Clearly identify infusion pumps used for epidural administration.
- Use KEYLOCK.

Table of Contents

Related product manuals