16
422348-Evolution ST350-Instruction manual-usa-rev01_AN
Blend mode (high-frequency energy combined with direct and direct current) is available for all hair removal middle,
coarse and very coarse options in preprogrammed mode.
You can select it by activating the «Blend» key and deactivate it to return to «Thermolysis» mode by activating the
«Thermolysis» key.
We advise you to harmoniously adjust (increase and decrease) the pre-recorded parameters (High-Frequency and
direct current) in the «Blend» mode.
We let you select this mode according to your own professionalism and judgment, knowing that it is more intended to
treat coarse, deformed, persistent hair deeply implanted in small areas.
On the treatment screen, the graph systematically represents the shape of the wave produced by the device. It is
specific to each zone, each technique and each size of hair.
IN BLEND MODE
• The pink arrows «Up / Down» in Electrolysis are used to increase or decrease at the same time the UL and the
intensity. Changing treatment parameters will affect potency.
• Time is not editable in Blend mode for High Frequency current (Thermolysis window). It was
calculated judiciously in order to produce the desired reaction of heating water molecules to
facilitate the caustic soda production
• The pink arrows «Up / Down» in Thermolysis are used to increase or decrease the intensity in %. Changing
treatment parameters will affect potency.