3 Vacuum system Sirona Dental Systems GmbH
3.5 Cleaning the suction hoses Care and cleaning instructions for the practice team C2
+
, M1
+
59 58 512 D 3370
10 D 3370.103.01.13.02
3.5 Cleaning the suction hoses
The hoses of the saliva ejector and suction cannula as
well as the connection hose to the water unit can be
pulled off from the assistant element for rinsing under
running water or thermodisinfection.
Switch the treatment center OFF!
Unscrew the suction hoses from the top parts (5) of
the filter housings.
Rinse the hoses under running water.
Thoroughly clean the top parts of the housings (5) with
the filter inserts every day.
The top parts of the filter housings can be thermodisin-
fected.
Remove the amalgam residue collectors (6) from the
lower parts (7) if the suction power decreases, otherwise
once a day. Store the amalgam residues under water in
a separate, closed container. When replacing the amal-
gam rotor (see page 12), put the residues into the rotor
for disposal.
ATTENTION
Do not dispose of amalgam residues in the cuspidor or
the sink!
When the surfaces of the hoses have become sticky
from disinfectants, clean them with a commercially avail-
able dish washing liquid and then powder them lightly
with talcum, if necessary.
Before refitting the hoses, grease the O-rings on the
connection pieces.
ATTENTION
Use only lubricants which have been approved by Sirona
(see page 2, General information)!
The hoses must engage firmly.
If the unit is equipped with a 3rd suction hose, clean it the
same way.
Grease