6 Disinfection system Sirona Dental Systems GmbH
6.3 Sanitation Care and cleaning instructions for the practice team C2
+
, M1
+
59 58 512 D 3370
20 D 3370.103.01.13.02
6.3 Sanitation
Sanitation intervals Perform sanitation every 4 weeks or whenever the
germination index significantly exceeds 100 germs
per ml (see section 6.1 “Microbiological water test”).
This is the only way to combat the formation of bio-
film in the water paths safely and effectively.
Special measures for additional devices Special devices that are connected to the chair via the
media block are treated with disinfected water with a
higher hydrogen peroxide concentration (1.4%) during
sanitation.
ATTENTION
If the connected devices are not suitable for this higher
sanitation concentration, Sirona recommends discon-
necting them from the treatment center during sanitation
and performing a separate sanitation in accordance with
the manufacturer’s specifications.
Preparations
Preparation for sanitation takes approx. 10 minutes.
•If DESINF flashes on the dentist element display,
first add 1 liter of DENTOSEPT P (see the Operating
Instructionsfor the , section on "Disinfection system
(optional)").
The flashing DESINF message then disappears.
Maximum water flow must be set on all burr drives
as well as on the SIROSONIC L. Check this before
every sanitation procedure.
All instruments and suction hoses must then be
placed in their holders.
A
DESINF
For reorders:
DENTOSEPT P 1 case = 6x1 liter,
Order No. 33 18 156
For Canada/Japan
only:
1 case = 6 x 1 liter,
Order No. 59 69 535