Sirona Dental Systems GmbH 3 Vacuum system
Care and cleaning instructions for the practice team C2
+
, M1
+
3.4 Suction handpieces
59 58 512 D 3370
D 3370.103.01.13.02
9
båÖäáëÜ
3.4 Suction handpieces
Sterilization
The following parts can be sterilized:
• the suction cannula (1)
• the intermediate piece (2)
• the suction handpiece (3)
• the saliva ejector handpiece (4)
• and the surgical suction cannula (5)
Sterilization only in the autoclave at 135°C (275°F), 2.1
bar, 3min.
Thermodisinfection
All parts can be thermodisinfected at 93×C (197°F).
Disinfection
All parts can be sprayed or wiped with a disinfecting
solution.
Use only cleaning agents which have been approved by
Sirona (see page 2, General information)!
Grease the O-rings at the joints afterwards!
Weekly
Separate the suction handpiece at the joint, clean it and
and grease the joints.
Use only lubricants which have been approved by Sirona
(see page 2, General information)!