Sirona Dental Systems GmbH 4 Amalgam separation
Care and cleaning instructions for the practice team C2
+
, M1
+
4.2 Cleaning the siphon
59 58 512 D 3370
D 3370.103.01.13.02
13
båÖäáëÜ
NOTE
i
Regardless of whether AMALG lights up on the display
or not, the rotor (3) in the amalgam separator must be re-
placed at least once a year. Otherwise unpleasant odors
can develop.
4.2 Cleaning the siphon
NOTE
i
Always wear work gloves when performing these tasks!
Clean the siphon (6) in case of poor drainage of the cus-
pidor, otherwise once a month.
Open cover flap (1) on base of water unit.
Unscrew the cap (5).
Have a collector ready to catch the contents!
Screw the cap back on and close the flap (1).