Sirona Dental Systems GmbH 3 Vacuum system
Care and cleaning instructions for the practice team C2
+
, M1
+
3.1 Care
59 58 512 D 3370
D 3370.103.01.13.02
7
båÖäáëÜ
3 Vacuum system
3.1 Care
After each patient
To ensure that the vacuum system is always ready for
use, you must aspirate a large glass of cold, clear water
through the suction hoses used after each patient,
especially after each intervention during which the
patient loses blood and after using hydrogen peroxide
(H
2
O
2
).
During prolonged treatments, you must aspirate a
glass of water at least every 60 minutes.
3.2 Disinfection with a dry suction system
Daily during the noon break and at the end of
each workday
The vacuum system is subjected daily to septic secre-
tions, saliva and blood. For hygienic reasons the system
must therefore be disinfected with a detergent/disinfec-
tant approved by Sirona.
ATTENTION
Use only cleaning and disinfecting agents which have
been approved by Sirona (see page 2, General informa-
tion)!
Do not use any washing powder or other disinfec-
tants!
Prepare the cleaning solution in a separate contain-
er according to the manufacturer’s instructions and
mix thoroughly.
Remove the gold trap.
Clean the cuspidor.
Plug in the cleaning adapter up to the stop. For the
saliva ejector hose, insert the intermediate piece (8)
into the cleaning adapter.
Fill the cleaning solution up to the filling limit (A) in
the cuspidor (approx. 0.75 liters).
Remove the suction hoses from their holders and at-
tach them to the side of the cleaning adapter, if pos-
sible all at once.
The cleaning solution is aspirated.
Observe manufacturer’s instructions regarding the
reaction time!