Sirona Dental Systems GmbH 8 Satelec Mini L.E.D. curing light
Care and cleaning instructions for the practice team C2
+
, M1
+
8.1 Care and cleaning of the handpiece and the connecting cable
59 58 512 D 3370
D 3370.103.01.13.02
29
båÖäáëÜ
8 Satelec Mini L.E.D. curing light
8.1 Care and cleaning of the handpiece and the connecting cable
The Mini L.E.D. and its accessories can be cleaned and
disinfected with cleaning and disinfecting agents that are
recommended by Sirona.
ATTENTION
Use only cleaning and disinfecting agents which have
been approved by Sirona (see page 2 , General informa-
tion)!
The connecting cable at the end of the handpiece can be
removed by turning it (see illustration below).
The Mini L.E.D. can be installed in holder 1 of the assis-
tant element and in holder 5 or 6 of the dentist element.
8.2 Care and cleaning of the light guide and the glare shield
Clean and sterilize the light guide and the glare shield
before every patient treatment (135 °C; 2bar; 3 min).
Check the light guide after each use.
There should be no traces of composite material on
the light guide.
Immediately remove any residue.
Make sure that the light guide is in perfect condition.
If you find any damage, immediately remove the light
guide, since damage will have a significant impact
on its performance.
Always use the glare shield to protect your eyes.
3 min
3 min