Sirona Dental Systems GmbH 7 Cleaning and care
Operating Instructions PerioScan
61 95 825 D 3496
D 3496.201.03.05.02
25
båÖäáëÜ
Foot switch
Clean or disinfect the bottom of the base plate using a
damp cloth moistened in water.
Do not pull on the connection cable when cleaning it.
Protection against medicaments
Due to their high concentrations and the substances
they contain, many medicaments can dissolve, etch,
bleach or discolor surfaces.
The only way to prevent damage is to wipe off medica-
ments immediately with a damp cloth!
Cleaning and disinfecting the rinsing media
tanks
The rinsing media tanks can be cleaned on the in-
side with hot water, a commercial dish washing liquid
and a bottle cleaning brush.
They are dish-washable and can also be thermally
disinfected in a thermo-disinfector at 95 °C (203 °F)
and 10 min. holding time.
ATTENTION
Perform thermal disinfection of the tanks only with their
covers removed.