Sirona Dental Systems GmbH 13 Electromagnetic compatibility
Operating Instructions PerioScan
61 95 825 D 3496
D 3496.201.03.05.02
37
båÖäáëÜ
13 Electromagnetic compatibility
13.1 Electromagnetic emission
The UNIT is intended for operation in the electromag-
netic environment specified below.
The customer or user of the UNIT should make sure that
it is used in such an environment.
The unit meets the requirements for electromagnetic
compatibility (EMC) in compliance with IEC 60601-1-2:
2001.
Emission measurement Conformity Electromagnetic environment guidelines
RF emissions according to
CISPR 11
Group 1 The UNIT uses RF energy only for its internal function. There-
fore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF emissions according to
CISPR 11
Class B The UNIT is intended for use in all facilities, including residen-
tial areas and in any facilities connected directly to a public
power supply providing electricity to buildings used for resi-
dential purposes.Harmonics
according to IEC 61000-3-2
not applicable
Power output
<50W
Voltage fluctuations/Flicker
according to IEC 61000-3-3
not applicable
no significant
flicker