48
WARNING: Using other manufacturer’s cables and accessories may affect EMC
performance. Unauthorized use of these items will void warranty and could result
in patient and/or user injury as well as damage to equipment or other property.
The use of cables or accessories other than those for which the bed was designed
or tested can significantly degrade emissions and immunity performance.
Cables and Accessories
Specifications and Requirements
Hand Control
Length: 9.51 feet (2.9 meters)
105°C (221°F)
24 AWG Strand
PVC Insulated
Standards: UL AWM Style 2464, NEC Type CMG/cUL CEC
Type CMG FT4, CSA AWM I A, EU CE
Power Cord
Hospital Grade
Length: 16 feet (4.88 meters)
SJT 16/3C 60°C (140°F)
End Types: YP-18L-2 and YC-12
UL and CSA certified 3 prong IEC C13 power cord receptacle.
Built to the IEC 60320-C13 standard. Rated 13A 125V/250V.
This IEC C13 power cord receptacle is fully molded with a low
profile ergonomic design and fully RoHS and REACH
compliant.
Coiled Cord (RSS)
US Plug 6P6C, Gold Plated 3u”
28 AWG (7/0.12BC)*6C+Jacket
Curve Length 1ft, 3ft L: (1ft*Æ13)+105+105mm
OD: 3.5mm
RoHS Compliant
Electromagnetic Interference (EMI)
The SW Evolutionand SW Rest Secure System™ have been tested and are intended for safe
use with other components compliant to IEC 601 standards for medical devices. The SW
Evolution and SW Rest Secure System™ can cause interference with non-601 regulated
equipment and/or other non-601 regulated equipment can cause interference with the SW
Evolutionand SW Rest Secure System™. This section is in accordance with IEC 60601-1-2
section 5.2.2.
WARNING: Radio wave sources such as radio stations, TV stations, amateur
radio (HAM) transmitters, two-way radios and cellular phones can affect this bed.
Following the warnings listed for electromagnetic interference should reduce the
chance of the bed’s unintended movement that could result in serious injury.
DO NOT turn ON hand-held personal communication devices, such as citizens band (CB) radios
and cellular phones while the bed is turned ON.