Guidance and Manufacturer’s Declaration – Immunity
All Equipment and Systems
The SW Evolution™ and SW Rest Secure System™ are intended for use in the electromagnetic
environment specified below. The customer, or user, of the SW Evolution™ and SW Rest
Secure System™ should ensure that it is used in such an environment.
Electromagnetic Environment
Guidance
Floors should be wood, concrete or
ceramic tile. If floors are synthetic, the
r/h should be at least 30%.
Mains power quality should be that of a
typical commercial or hospital
environment.
±1kV
Differential
±2kV
Common
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage Dips/Dropout
IEC 61000-4-11
>95% Dip
for
0.5 Cycle
60% Dip for
5 Cycles
30% Dip for
25 Cycles
>95% Dip
for
5 seconds
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the bed
requires continued operation during
power mains interruptions, it is
recommended that the bed be powered
from an uninterruptable power supply or
battery.
Power Frequency
50-60Hz
Magnetic Field
IEC 61000-4-8
Power frequency magnetic fields should
be that of a typical commercial or
hospital environment.