Page 9
WARNING
The mounting plate must be accurately
leveledwithin0.1°topreventlighthead
drift.
NOTE
Specific seismic calculations are avail-
able if required.
2. Radial Arm Assembly and Ceiling Cover
NOTE
•Themultiplearmassembliesareeasier
to handle during installation if the arms
are left taped and tied together.
•In some cases it maybe necessary
to connect the electrical wires from the
wall control to the radial arm junction
box before the arm assembly can be
bolted to the mounting plate.
•Usetwo6"to8"boltstoholdxture
while connecting wiring.
a. Install the Radial Arm Assembly(RAA) onto
the mounting plate using the bolts provided. Tighten
the mounting bolts securely.
NOTE
Radial Arm wires are tagged for proper
connection to the Wall Control (top arm
#1, next arm #2, bottom arm [triple arm
models] #3).
Figure 3. Radial Arm Installation
Figure 2. Mounting Plate Installation
b. Observe wire tags and color codes and con-
nect the electrical wires from the wall control to the
radial arm junction box wires.
CAUTION
Connection of the fixture wires must
be made using crimp connectors. Main
terminal devices shall be so located or
shielded that, should a wire of a stranded
conductor escape when the condutors
are fitted, there is no risk of accidental
contact between live parts and ac-
cessible parts. Acceptable sheilding
methods include UL approved shrink
tubing and electrical tape.
c. Install the ceiling cover and secure. See
figure 3.
FINISHED
CEILING
3/4"
SUPPORT ROD
SWAY
BRACING
STIFFENER PLATE
JAM NUTS
MOUNTING PLATE
1-3/8"
FINISHED CEILING
RADIAL ARM
CEILING
COVER
MOUNTING
PLATE