CONDUIT REQUIREMENTS
This fixture requires 2 dedicated conduit raceways at the wall control to separate the 120VAC supply
lines from the 30VDC supply lines to the lighting fixture. Failure to observe this requirement will allow
the migration of Electrical Magnetic Interference and will disrupt the operation of the lights.
Use of approved metal conduit shall be employed throughout the fixture's wiring circuit where appli-
cable.
PROTECTIVE MEANS
This fixture requires a properly circuit protected, dedicated, 120VAC, 15 amp circuit. An isolated
power supply circuit must be protected by a 15 amp, double pole, single throw circuit breaker.
Proper performance and safety of this fixture can only be achieved by an adequate grounding sys-
tem. Fixture ground must be a dedicated ground point ultimately bonded to the facilities grounding
system to prevent the migration of electrical interference generated by other devices.