WARNING
The antistatic properties of the table are
dependent on the use of the original pad
set which was furnished with the table or
an alternate approved replacement.
WARNING
Certain accessories may limit weight
capacities. Check with your SKYTRON
representative.
NOTE
Always follow current AORN Journal
Guidelines to ensure proper cleaning
and disinfection procedure.
WARNING
Always follow OSHA blood-borne patho-
gens standards for protective clothing,
including gloves, masks and eye pro-
tection when cleaning the surgical table.
CAUTION
Thoroughly read and follow the
manufacturer's directions for all clean-
ing fluids. DO NOT use cleaners con-
taining phenolics.
CAUTION
When using spray cleaners DO NOT
spray fluids directly into electrical re-
ceptacles or micro switches.
CAUTION
Before replacing pads on the table, make
sure the pads and all mating surfaces
are completely dry. Moisture trapped
between the pads and mating surfaces
may cause distortion of table tops.
NOTE
•When the red light starts to blink (indi-
cating low power in battery) the table
will operate for approximately 5 con-
tinuous minutes, typically long enough
to use the table for the rest of the day.
•The charging system operates ONLY
when the table is in AC 120V mode.
NOTE
In battery mode, only the red indicator
light will be on. In AC 120V mode, both
the green indicator light and red indica-
tor light will be on.
NOTE
The table can be operated on 120VAC
power while the battery is being re-
charged.
NOTE
The emergency back-up control
switches will function when the table is
operating on 120VAC power, battery
power, or turned off.
CAUTION
•The Emergency Release Valve must
be closed and tightened (clockwise)
before activating any hydraulic func-
tion.
•If the Emergency Brake Release Valve
has been operated, the UNLOCK but-
ton on the pendant control will have to
be pressed before brakes will lock
again.
WARNING
Consult manufacturer's instructions
when using high frequency surgical
equipment, cardiac defibrillator and car-
diac defibrillator monitors.
WARNING
When an antistatic pathway is required,
the table has to be used on an antistatic
floor.