SECTION IV MAINTENANCE
4-1. Preventive Maintenance
The following preventive maintenance checks and
services are recommended to ensure the service-
ability and proper operation of your SKYTRON
Surgical Table.
a. During normal cleaning, a general visual ex-
amination should be made checking for leaks,
loose bolts or parts and cracked, chipped, or miss-
ing paint. Any necessary repairs should be made.
b. Semi-annually the following checks and ser-
vices should be performed:
1. Check all hydraulic fittings, mini-valves
and slave cylinders for proper operation
and any signs of leaks.
2. Check the hydraulic speed controls and
adjust if necessary.
3. Pressure check (with a gauge) the pres-
sure relief valve.
4. Check all mechanical adjustments and
adjust as necessary.
5. Check hydraulic fluid level.
6. Lubricate the elevation slider assembly.
7. Lubricate the top slide components.
4-2. Cleaning Recommendations
NOTE
Always follow current AORN Journal
Guidelines to ensure proper cleaning
and disinfection procedure.
The following procedures should be followed when
cleaning the surgical table between cases.
Place table top in level position prior to starting
cleaning procedure.
WARNING
Always follow OSHA blood-borne patho-
gens standards for protective clothing,
including gloves, masks and eye pro-
tection when cleaning the surgical table.
Remove major contaminants from the table with
disposable materials following appropriate biohaz-
ard waste disposal procedures.
Remove all table pads and place them on a flat
surface for cleaning.
CAUTION
Thoroughly read and follow the
manufacturer's directions for all clean-
ing fluids. DO NOT use cleaners con-
taining phenolics.
Apply cleaning fluid liberally to top and sides of
each pad and wipe with a clean lint-free cloth.
Using a clean, damp, lint-free cloth, wipe the pads
to remove the cleaning fluid.
Using a clean, dry, lint-free cloth, wipe the pads to
remove all moisture.
Repeat the steps to clean the bottom of the each
pad.
CAUTION
When using spray cleaners DO NOT
spray fluids directly into electrical re-
ceptacles or micro switches.
Repeat cleaning procedure for all table surfaces
including the top, sides, elevation column, base
and all accessories.
CAUTION
Before replacing pads on the table, make
sure the pads and all mating surfaces
are completely dry. Moisture trapped
between the pads and mating surfaces
may cause distortion of table tops.

Other manuals for Skytron ELITE Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Skytron ELITE Series and is the answer not in the manual?

Skytron ELITE Series Specifications

General IconGeneral
CategoryMedical Equipment
Light SourceLED
Lux Intensity160, 000 Lux
Diaphragm ControlYes
Shadow ManagementYes
Ceiling MountYes
Warranty5 years
TypeSurgical Light
SeriesELITE

Related product manuals