151
7.3 Instruction Set for RS-Communication
The commands consist of three parts: Address two-digit aa, e.g.: 01
Command e.g.: DA
Variable, if necessary e.g.: 14
and end of command <CR> <LF>
Every command must be completed with the ASCII - sign <CR> and <LF> (Carriage Return and Line
Feed). Only if the respective action has ended the answers will be returned to the computer.
Example: The command to dose 12.5 ml shall be sent to the burette TitroLine® 5000 with the
address 2.
The command consists of the characters: 02DA12.5<CR LF>
In detail: 02 = Device address
DA = Dosage command with filling and zero points of the display
12.5 = Volume in ml to be dosed
<CR LF> = Control character as command end
Command Description Reply
aaAA automatic allocation of device address aaY
aaMC1...XX choosing a method aaY
aaBF “filling burette“. Aufsatz wird gefüllt. aaY
aaBV output of dosed volume in ml aa0.200
aaDA dose volume without filling, with adding the volume aaY
aaDB dose volume without filling, reset of the volume aaY
aaDO dose volume with filling, without adding the volume aaY
aaGF filling time in seconds (min is 20, default 30) aaY
aaEX “exit“ function.back to main menu aay
aaFP pH measurement function aay
aaFT temperature measurement function aay
aaFV mV measurement function aay
aaGDM dosing speed in ml/min (0.01 100 ml/min) aaY
aaGF filling time in sec (adjustable 20 999 seconds) aaY
aaGS output serial no. Of device aaGS08154711
aaLC output of the CAL parameters
aaLD output of the measurement data aaY
aaLR output report (short report) aaY
aaM output of the preset measurement value (pH/mV/ug) aaM7.000
aaLI output method content
aaRH request of identification aaIdent:TL500
aaRC send last command aa“last command“
aaRS report status aaStatus:“text
possible answers are:
”STATUS:READY“ for ready
”STATUS:dosing“ dosing
”STATUS:filling“ filling
”ERROR:busy“ if no interchangeable unit has been attached
aaSM start selected method aaY
aaSEEPROM EEPROM reset to factory defaults aaY
aaSR stop the actual function aaY
aaSS titration start with the transfer of the pH end value aaY
aaVE Version number of the software aaVersion