1.2 Specifications Titrator TitroLine
®
5000
State 01.09.2014
CE sign:
EMC compatibility according to the Council Directive: 2004/108/EG;
applied harmonized standards: EN 61326-1:2006
Low-voltage directive according to the Council Directive 2006/95/EG
Testing basis EN 61 010, Part 1
Country of origin: Germany, Made in Germany
The following solvents/titration reagents are allowed to be used:
All common titration solutions.
As reagent water and all non-aggressive non-organic and organic fluids are allowed. If using combustible
fluids fire please adhere to the Guidelines for Explosion Protection and Prevention of the chemical industry.
For fluids with higher viscosity ( 5 mm
2
/s), lower boiling point or affinity to outgas, the filling and dosage
speed can be adjusted.
Fluids with viscosity over 10mm
2
/s cannot be dosed.
Measuring input: pH/mV-input with 12 bit transducer for high-precision readings.
Electrode socket according to DIN 19 262 or
additional with BNC socket insert (Z 860)
Reference electrode 1 x 4 mm socket
Measurement range
Display
resolution
Measurement
accuracy* without
sensor probe
Input
resistance

pH
- 3.0 ... 17.00
0.01
0,05 1 Digit
> 5 10
12
U mV
- 1900 ... 1900
1
1,0 1 Digit
> 5 10
12
To ensure maximum accuracy of the readings we recommend to allow some reasonable time for the
TitroLine® 5000 titrator to “warm up”.
Measurement input: Temperature probe - connector for a Pt 1000 resistance thermometer and NTC 30
Connection: 2 x 4 mm - sockets and 1 x 2 mm socket.
Measurement
range
Display resolution
Measurement
accuracy* without
Pt 1000 sensor
Measurement
accuracy* without
NTC 30 sensor
T [°C]
- 30 ... 115
0.1
0.5 K 1 Digit
0.5 K 1 Digit
* The measurement uncertainty of the sensor probe has to be taken into account as well
Display: 3.5 inches -1/4 VGA TFT display with 320x240 pixels.
Calibration: Automatically with up to three buffer solutions, sequence during calibration optional,
freely definable buffers can be input.
Default buffer solutions according to DIN 19 266 and NBS, or technical buffers:
pH = 1.00; pH=4.00; pH=4.01; pH=6.87; pH= 7.00; pH= 9.18; pH=10.00;
Inputs:
Input pH/mV: pH/mV-input with electrode socket according DIN 19 262/or BNC
Input Pt 1000/NTC 30: Temperature sensor probe (Connection sockets: 2 x 4 mm
Power supply: power supply 100 240 V; 50 60 Hz, power input: 30 VA
Use the Power supply TZ 1853, Type No.: FW 7362M/12 only!