Other Settings
●Peak/Bottom value display
The maximum (minimum) instantaneous flow, from when the power was supplied to this moment, is
detected and updated.
In the peak/(bottom) value display mode, the maximum (minimum) instantaneous flow can be displayed.
For peak value display mode, when the button is pressed for 1 second or longer, the maximum flow will
be displayed flashing, and is held.
To release holding the peak value display, press the button for 1 second or longer to return to
measurement mode.
For bottom value display mode, when the button is pressed for 1 second or longer, the minimum flow
will be displayed flashing, and is held.
To release holding the bottom value display, press the button for 1 second or longer to return to
measurement mode.
If the and buttons are pressed simultaneously for 1 second or longer while the flow is being held,
the peak (bottom) value is reset.
●Zero Clear
The displayed value can be adjusted to "0" when the measured flow is within ±10% F.S. of the zero point at
the time of shipment from the factory.
(A tolerance of ±1 digit may apply due to individual product differences)
Press the and buttons simultaneously for 1 second or longer, to clear the display to "0". The display
will return to measurement mode automatically.
The accumulated value is "0" when accumulated flow is displayed.