MIC-5010, MIC-5005 USER MANUAL
38
e) dimensions ......................................................................................... 390 mm x 310 mm x 170 mm
f) meter weight ................................... approx. 5.6 kg with Li-Ion battery; approx. 7 kg with gel battery
g) storage temperature ................................................................................................. 25C...+70C
h) working temperature .................................................................................................. -20C...+50C
i) humidity .......................................................................................................................... 20%...90%
j) altitude (above sea level) ................................................................................................... ≤3000 m
k) reference temperature ................................................................................................. +23C ± 2C
l) reference humidity .......................................................................................................... 40%...60%
m) display ........................................................................................................................ modular LCD
n) number of R
ISO
measurements according to EN 61557-2 .................................................. min. 1000
o) memory of measurement results ........................................................................................ 990 cells
p) data transmission ...................................................... USB connection or wireless by OR-1 receiver
q) quality standard ............................ design, construction and manufacturing are ISO 9001 compliant
r) the device meets the requirements of the EN 61010-1 and IEC 61557 standards
s) the product meets thr EMC requirements (immunity for industrial environment) according to the fol-
lowing standards ............................................................ EN 61326-1:2006 and EN 61326-2-2:2006
ATTENTION!
MIC-5010 and MIC-5005 meters are classified in terms of Electromagnetic Com-
patibility (EMC) as instruments of Class A (for use in industrial environments -
according to EN 50011). Interferences, impacting the operation of other de-
vices must be taken into account when the meters are used in other environ-
ments (e.g. domestic).
11.2 Additional data
Data on additional uncertainties are useful mainly when the meter is used in non-standard condi-
tions and for metrological laboratories for the purpose of calibration.
11.2.1 Additional uncertainties according to IEC 61557-2 (R
ISO
)
Significant parameter
Designation
Additional uncertainty
Position
E
1
0 %
Supply voltage
E
2
2% (BATis not lit)
Temperature 0C...35°C
E
3
6 %
11.2.2 Additional uncertainties according to IEC 61557-4 (R ±200mA)
Significant parameter
Designation
Additional uncertainty
Position
E
1
0 %
Supply voltage
E
2
0.2% (BATis not lit)
Temperature 0...35°C
E
3
1 %