EasyManua.ls Logo

Sonoscape A6T

Sonoscape A6T
172 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-78
Measurement method:
1. In B mode freeze the image, press the Calckey to pop up the
Meas & Calc menu, move the trackball, use the cursor to select
“Cardiac”, press theSETkey to go to the submenu, move the
trackball to select “LVM”, press theSETkey to go to measurement
state.
2. Measure LVPWd, see “Distance Measurement” for the specific
method.
3. Measure IVSTd, see “Distance Measurement” for the specific
method.
4. Measure LVIDd, see “Distance Measurement” for the specific
method.
5. Based on the formula to calculate LV Mass, the measurement and
calculation result is displayed in the result window.
6. Repeat steps 1 to 5 to go to the next LV Mass measurement.
3.6.2.4.5 Left Ventricle Mass Weight Index (LVMWI)
Function: Based on patient’s height, weight and LV Mass, to calculate
LVMWI.
The formula:
LVMWI=LV Mass (g)/BSA (m
2
)

Table of Contents

Related product manuals