A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-76
3.6.2.4.2 Right Ventricle inner diameter (RV)
Function: To measure Right Ventricle inner diameter.
Measurement method:
8. Measure LVAPs, see “Area/Cir measurement
(Ellipse)” for the specific measurement
method.
9. Based on the formula you can get EDV, ESV,
SV, CO, EF, SI, and CI. The measurement and
calculation result is displayed in the result
window.
10. Repeat steps (1) to (9) to perform the next
measurement.
1. In B mode freeze the image, press the
Calckey to pop up the Meas & Calc
menu, move the trackball, use the cursor to
select “Cardiac”, press theSETkey to go to
the submenu, move the trackball to select
“RV”, press theSETkey to go to
measurement state.
2. Measure RV; see “Distance Measurement”
for the specific method.
3. The measurement result is displayed in the
result window. Repeat steps 1 to 2 to
perform the next measurement.