3.6.2.3.3 L.Ovary (Left Ovary) measurement
Function: Based on left ovary length, height and width: Lt L, Lt H, Lt W,
to calculate left ovary volume: L.OV-V = 0.523 * Lt Lmm * Lt Hmm *
LtWmm
Measurement method:
1. Press the Calckey to pop up the Meas &
Calc menu, move the trackball, use the cursor
to select “GYN”, press theSETkey to go to
the submenu, move the trackball to select
“L.Ovary”, press theSETkey to go to
“L.Ovary” measurement mode.
2. Measure Lt L of left ovary, sees “Distance
Measurement” for the specific method.
3. Measure Lt H of left ovary, sees “Distance
Measurement” for the specific method.
4. Measure Lt W of left ovary, sees “Distance
Measurement” for the specific method.
5. L.Ovary volume L.OV-V = 0.523 * Lt Lmm * Lt
Hmm * Lt Wmm, and the measurement &
calculation result is displayed in the result
window.
6. Repeat steps 1 to 5 to go to the next L.Ovary
volume measurement.