A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
D-2
Table 1-2. Transducer Model: 6-2MHz C351 Curved Array
Operating Mode: THI-mode
TIS
Non-scan
Index
label
MI
Scan
A
aprt
1c A
aprt
>1cm
TIB
Non-
scan
TIC
Maximum index value
0.54 0.11
# # # #
p
r, α
(MPa)
1.02
P (mW)
6.3
# # #
min of (mW)
#
z
s
(cm)
#
z
bp
(cm)
#
z
b
(cm) 4.7 #
z at max I
pi.α
(cm) 4.7 #
d
eq
(z
b
) (cm) #
f
awf
(MHz)
3.6 3.6
# # # #
Dim of A
aprt
X (cm) 1.9 # # # #
Associated
acoustic
parameters
Y (cm)
1.3 # # # #
t
d
(µsec)
0.476
prr (Hz)
4219
p
r
at max I
pi
(MPa) 1.82
d
eq
at max I
pi
(cm) #
I
pi.α
at max MI (W/cm
2
) 25.6
FL
x
(cm) 5 # # #
Other
information
Focal
Length
FL
y
5 # #
#
Mode
THI THI
# # #
#
Focus
(cm)
5 5
# # #
#
PRF
(kHz)
4 4
# # # #
Operating
control
conditions
Power
(%)
100 100
# # #
#