A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
D-6
Table 2-3. Transducer Model: 6-2MHz C352 Curved Array
Operating Mode: M-mode
TIS TIB
Non-scan
Index label MI
Scan
A
aprt
1c A
aprt
>1cm
Non-
scan
TIC
Maximum index value
0.65 #
# 0.21 0.58 #
p
r, α
(MPa) 1.14
P (mW) # # 32.5 #
min of (mW)
15.2
z
s
(cm) 4.6
z
bp
(cm) 2.65
z
b
(cm) 4.5 4.5
z at max I
pi.α
(cm) 4.5 4.5
d
eq
(z
b
) (cm)
0.04
f
awf
(MHz) 3.1 # # 3.1 3.1 #
Dim of A
aprt
X (cm) # #
1.9 1.9
#
Associated
acoustic
parameters
Y (cm)
# # 1.4 1.4 #
t
d
(µsec) 0.553
prr (Hz) 4219
p
r
at max I
pi
(MPa) 1.82
d
eq
at max I
pi
(cm) 0.04
I
pi.α
at max MI (W/cm
2
) 30.1
FL
x
(cm)
# # 5 #
Other
information
Focal
Length
FL
y
(cm
# # 5
#
Mode
M #
#
M M #
Focus
5 #
#
5 5 #
PRF
4 #
#
4 4 #
Operating
control
conditions
Power 100
#
# 100 100
#

Table of Contents