What to do if limit values are not maintained during iCALising SPECTRO SPECTROMAXx?
- LLinda SmithAug 5, 2025
If the limit values are not maintained during iCALising of your SPECTRO Measuring Instruments, it could be due to insufficient intensity. To address this, first, check the argon supply and sample. Then, clean the spark stand and adjust the electrode. Next, clean the light inlet window. Finally, ICALize the instrument again. If the issue persists, service center assistance may be required. Alternatively, high intensities may be the cause. In this case, check the shutter function and the plasma generator parameters.