Operation
28 SPECTROMAXx
6.3 Switching off the instrument
Information
If the instrument was completely switched off, restarting may take some time as
the optics need to reach the right temperature and the system must be flushed with
argon.
It is recommended to leave the instrument connected to the power supply and to
leave the argon supply line open all the time.
To switch off the instrument, proceed as follows:
1. Turn off the plasma generator by pressing the switch.
2. Exit the spectrometer program "Spark Analyzer Pro MAXx".
3. Select the dialog item "Shut down" and confirm the dialog field.
The program is shut down.
The computer is turned off.
4. In order to switch the instrument completely off, remove the power plug.
The instrument is switched off.