16 SPECTROMAXx —
2.2.4 Sample weight
The weight of the sample to be measured must not exceed 15 kg (33 lbs).
2.2.5 Spark stand
The opening of the spark stand plate must be completely covered by the sample
or the respective adapter.
Warning! Optical radiation!
The plasma in the spark stand
generates UV radiation. UV radiation may cause
damage to the eyes or skin.
Use the appropriate sample adapter for samples that are not completely
covered by the spark stand opening.
Take appropriate protective measures (e.g. wear safety goggles) when the
spark stand opening is not fully covered.
2.2.6 Sample clamp
The sample clamp must have contact with the sample (the safety device is
activated).
Warning of hazardous voltage!
The sample clamp provides the contact between the sample and the spark
stand.
There is a risk of suffering an electric shock when handling the sample clamp.
Never tamper with the sample clamp.
Warning of hand injury!
Springs are used to keep the sample clamp in position. However, the clamp may
fold down
unexpectedly. If your hand is caught underneath the clamp, this may
Never place your hand under the sample clamp.