Technical data
20 SPECTROMAXx
3 Technical data
Description Value
Instrument height Benchtop instrument: 450 mm; 17.7"
Floor instrument: 1350 mm; 53.1"
Instrument width Benchtop instrument: 625 mm; 24.6"
Floor instrument: 625 mm; 24.6"
Instrument depth Benchtop instrument: 750 mm; 29.5"
Floor instrument: 790 mm; 31.1"
Instrument weight Benchtop instrument: approx. 78 kg; ~ 172 lbs
Floor instrument: approx. 132 kg; ~ 291 lbs
Power supply 90 - 253 VAC, 50/60 Hz
Power during measurement 400 VA
Power during stand-by 180 VA
Supply voltage fuse 5 A (slow-blow)
Operating range (permissible ambient
temperature)
10°C 30°C; 50°F 86°F
Recommended temperature range to
achieve the specified performance
18°C 28 °C; 64°F 82°F
Place of installation Inside of dry, dust-free rooms up to an altitude of
2000 m above sea level, vibration-free (not
subject to impact or vibration load conditions)
Air humidity range 20 80%, not condensing
Sound pressure level 55 dB(A)