Technical data
SPECTROMAXx
21.02.2020
21
Description Value
Argon inlet pressure on the instrument
4.8 to 5.0 bar (69.6 to 72.5 psi)
Argon quality (min.) 4.8 (99.998% Ar) with O
2
<3ppm, N
2
<10ppm,
H
2
O<5ppm, kV<1ppm
Argon of higher purity or gas pre-purification is
required for a nitrogen analysis.
Argon flow rates LMM08 LMF08/LMD08
Constant flow
Analytic flow
25 l/h
227 l/h
11 l/h
213 l/h
Storage conditions Ambient temperature
-15 45°C; +5 +113°F
Protection class / Contamination class
IP 20 (DIN 40050), contamination class II
(laboratory environment)
EMC environment Industrial environment / Laboratory environment