Operation
SPECTROMAXx
21.02.2020
31
iCALising the measuring device (iCAL)
The iCALising procedure (standardizing using iCal) requires a reference sample.
The average values of these measurement results are calculated and stored in the
instrument.
In order to proceed with iCALization, follow the steps below:
1. Place a clean electrode in the spark stand, see page Fehler! Textmarke
nicht definiert. ff.
2. Start the iCAL function by pressing the the “iCAL” button.
3. Follow the instructions on the screen and perform the necessary
measurements with the reference sample.
4. Poor measurements are highlighted in the ICAL measurement window and
can be deleted. Please repeat the respective measurements.
The instrument is iCALized.
Changing the base (multi-base instruments only)
When changing the base, the spark stand must be cleaned in order to avoid
memory effects, see page 40.
Change the spark stand insert, the electrode, and the electrode brush.
Preparation of the instrument for measuring another base is now complete.