Maintenance
SPECTROMAXx
21.02.2020
37
7.4 Cleaning the UV optics light inlet window
Warning of hazardous voltage!
When the plasma generator is switched on, there is a risk of electric shock if the
electrode is touched.
Shut down the plasma generator when working on the spark stand.
Warning! Toxic substances!
When
removing the light entry window, contact with the condensate of the
samples is likely. Depending on the sample material, the condensate may be toxic.
Protective gloves must be worn.
To clean the light inlet window, proceed as follows:
1. Turn off the Plasma generator
(1).
2. Unscrew the 4 fasteners (2) and
remove the spark stand plate
(3).
Fig. 12
3. Remove the spacer (1) from the
spark stand.