Use only original or Spencer Italia S.r.l. approved components/replacement parts and/or accessories to carry out any operation without causing alterations or modifications
to the device. Otherwise, we decline all responsibility regarding incorrect operation or any damage caused by the device to the patient or the operator, invalidating the
warranty and invalidating compliance with EU Regulation 2017/745.
12.3 PERIODIC OVERHAUL
No periodic overhaul is foreseen for the device.
12.4 SPECIAL MAINTENANCE
Special maintenance can only be carried out by the Manufacturer, who uses internal and external technicians specialised and authorised by the Manufacturer itself.
Only maintenance activities carried out by specialised technicians authorised by the Manufacturer are considered valid by Spencer Italia S.r.l..
The end-user can replace only the spare parts indicated in § 15.
12.5 LIFE SPAN
The device, if used as described in the following instructions, has a life span of 10 years from the date of purchase.
Belts shall be replaced every 2 years.
Spencer Italia S.r.l. will accept no responsibility for incorrect operation or damage caused by the use of devices that have exceeded the maximum allowable life span.
13. TROUBLESHOOTING TABLE
PROBLEM CAUSE REMEDY
Damage to the shell, to riveting and/or the stretcher structure Improper use
Immediately remove the stretcher from service
and contact the Customer Care Service
Difficulties in assembling parts (Twin Shell only)
Dirty inlets
Clean inlets thoroughly in both the convex and the
concave parts
Change shells, if necessary
Make sure that there has not been an exchange
of other Twin Shell shells that may be in your
possession
Damaged pins and/or linchpins
Immediately remove the stretcher from service
and contact the Customer Care Service
The stretcher does not stay aligned when lifted (Twin Shell only) Parts assembly error
Dismantle and reassemble the parts, carefully
checking the stretcher
The stretcher does not stay joined (Twin Shell only) Safety device and/or linchpin broken
Immediately remove the stretcher from service
and contact the manufacturer
Breakage of the patient support bed and/or integrated spine board Improper use
Immediately remove the stretcher from service
and contact the manufacturer
The back lift piston and/or the Trendelenburg position are not
locked/in position (Boston Tec and Boston Light model only)
Damaged locking mechanism
Immediately remove the stretcher from service
and contact the manufacturer
Spine board release mechanism does not work (Boston Pro model
only)
Possible dirt inside the mechanism Thoroughly clean the mechanism
Breakage of the lock/release mechanism
Immediately remove the stretcher from service
and contact the manufacturer
The perimeter rope is not sufficiently taut (Shell, Twin Shell and
Dakar models only)
The rope may be worn or no longer atta-
ched to the main structure
Immediately remove the stretcher from service
and contact the Customer Care Service
Failed blocking of telescopic handles (Dakar model only) Breakage of the internal locking mechanism
Immediately remove the stretcher from service
and contact the Customer Care Service
Wheels do not slide (Dakar model only)
Excessive wear on wheels
Immediately remove the stretcher from service
and contact the Customer Care Service
Damaged wheel support or wheel itself
Immediately remove the stretcher from service
and contact the Customer Care Service
The footrest is not stable after fixing to the stretcher (Shell, Twin
Shell , Dakota , Dakota Light and Dakar models only)
Damage to buckle structure or to the
structure of the footrest or adjustment belt
Immediately remove the stretcher from service
and contact the Customer Care Service
Belts do not stay joined Damaged buckle mechanism
Immediately remove the stretcher from service
and contact the Customer Care Service
If a problem or fault is detected that does not correspond to the above, please contact Spencer Italia srl customer care service.
14. ACCESSORIES
14.1 ACCESSORIES
CODE DESCRIPTION COMPATIBLE
ST00592A STX 598 - 2 PIECE YELLOW BELT W/METAL BUCKLE All versions
ST70002A STX 702 - TWO PIECE METAL BLACK REFLEX BELT All versions
ST04519C STX 519 - ADJUSTABLE HARNESS SYSTEM SHELL / TWIN-SHELL / DAKAR / BOSTON TEC / BOSTON LIGHT / BOSTON PRO
ST04522B STX 540 - FIXED HARNESS SYSTEM SHELL / TWIN-SHELL / DAKAR / BOSTON TEC / BOSTON LIGHT / BOSTON PRO
ST04518A STX 518 - 2-PART UNIVERSAL FLOAT FOR BASKET SHELL / TWIN-SHELL / BOSTON TEC / BOSTON LIGHT / BOSTON PRO
ST04524B STX 538 - 3-PART UNIVERSAL FLOAT FOR BASKET SHELL / TWIN-SHELL / BOSTON TEC / BOSTON LIGHT / BOSTON PRO
ST04040A STX 40 - YELLOW BAG FOR SHELL SHELL / BOSTON TEC / BOSTON LIGHT / DAKOTA / DAKOTA LIGHT
ST04110A STX10 - YELLOW BAG FOR TWIN SHELL TWIN-SHELL
15. RICAMBI
CODE DESCRIPTION FOR MODEL
ST00592 STX 592 - 2 PC YELLOW BELT METAL BUCKLE All
ST70002 STX 702 2pcs BLACK REFLEX BELTS w/METAL BUCKLE, All
RIST111 MATTRESS FOR BASKET STRETCHERS WITH ADHESIVE Shell/Twin Shell

Table of Contents

Other manuals for Spencer SHELL

Related product manuals