17
5. Specific Operation Manual
5. Specific Operation Manual
left/right (◀▶) Keys.
* Turbidity changes in the order of CLEAR, SL CLOUDY,
CLOUDY, TURBID and OTHER, pressing left/right (◀▶)
Keys.- If patient’s/sample’s IDs are serial numbers, you may
register - If
- If patient’s/sample’s IDs are serial numbers, you may register
only the first patient’s/sample’s ID. Then the others will be
numbered in sequence, automatically.
- If patient’s/sample’s ID includes letters, you should enter IDs
one by one even though IDs are serial numbers.
- In case you don’t register patient’s/sample’s ID before a test, ID
columns remain blank.
5.8 Measurement
1) A measurement should be started in Quick Mode or Normal
Model during Strip Loading Plate is out of the unit.
2) Dip a test strip into fresh urine sample for a few seconds and take
it out.
3) Remove excessive urine by contacting the side of strip on the
material such as soft tissue.
4) Lay a strip, there is buzz sound. After 3 seconds loading plate will
enter the unit.
5) In Normal Mode, result will be shown on the screen and LCD
after 100 seconds. In Quick Mode, result will be shown
immediately.
6) Analyzer will stand by after the measurement. If there is no strip
to be tested on the strip loading plate, the UroMeter 120 stops
working.
7) Whenever press ENT, STAND BY MODE ▶ Quick Mode ▶
Normal Mode, Strp loading plate will move relatedly.
Precautions ;
1) You must use strips of SD UroColor 4~11 series manufactured by
Standard Diagnostics, Inc.
2) Do not touch the strip pad by hand.