22
6. How to store after using
6. How to store after using
6.1 Daily Check Up
Separate Strip loading plate from the instrument and wash them in
water.
6.2 Cautions for Keeping
1) Avoid the places subject to high temperature, high humidity,
direct sunlight and air containing dusts, salt and sulfur contents.
2) Avoid a slope or vibrating place for keeping.
3) Avoid any impact when instrument is moved.
6.3 Error, errors!
A bent strip will cause a jam inside of Strip loading plate of
instrument. In this case, you take out the front cover and remove the
strips jammed.
6.4 Error Messages and Troubleshooting
“Misplaced
strip”
Cause : No test strip is sensed by sensor or the
strip is incorrectly positioned on the sensor.
Action : Repeat the measurement with new strip
Cause : When you turn on the power switch,
internal calibration will be performed
automatically. In this process, if the strip
loading plate is incorrectly inserted or out
of working normally, this error message
will be shown.
Action : Please turn off and restart the machine. If
it happen repeatedly, please contact your
local technical Service.
“ERR Load
Over Cnt”