NOTE: Be sure that system is not delivering RF energy while changing probes.
1. Twist and pull connector on Handpiece Cable to detach from Probe. When Probe is fully detached, the CUT
Level window will display a ‘0’.
2. Properly dispose of the used probe. Probes are for single use only. DO NOT attempt to reuse.
3. Attach new Probe to the Handpiece Cable. When new Probe is attached, the default setting for that particular Probe
will be displayed in the CUT Level window and a beep will sound. If using the Synthesized Voice feature, the default
CUT Level will be announced following proper connection.
WARNING: Do not insert or withdraw Probe while activated. Injury to patient and/or damage to the equipment may
result.
NOTE: If Generator is turned off while Probe is still attached or there is a loss of power to the unit, the Generator will
automatically return to the default output level when power is restored. Power can be increased/decreased
using the arrow buttons on Generator or the plus/minus pedals on the Footswitch.
RECOMMENDRECOMMEND
RECOMMENDRECOMMEND
RECOMMEND
AA
AA
A
TIONS DURING SURTIONS DURING SUR
TIONS DURING SURTIONS DURING SUR
TIONS DURING SUR
GERGER
GERGER
GER
YY
YY
Y
::
::
:
The output power selected should be as low as possible for the intended purpose to prevent injury to patient.
Avoid prolonged activation of the Probe. The system allows for a maximum of 160 seconds. If activation exceeds
160 seconds, an error alarm will sound and an error code will be displayed in the CUT Level window.