Stryker SERFAS - Setting Up the Probe

Stryker SERFAS
35 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SERFAS Probe
Overall Length .......................................................................................... 8.7-13.2in (22 – 33.5cm)
Working Length ....................................................................................... 3.93- 6.9in (10 – 17.5cm)
Shaft Diameter .................................................................................... 0.06 – 0.18in (1.5 – 4.5mm)
Shaft Bend Angle .................................................................................................................. 0 – 45º
Active Tip Orientation ........................................................................................................... 0 – 90º
Supplied Sterile and For Single-use ONLY
Sterilization Method ......................................................................................................... Irradiation
SERFAS Handpiece Cable
Overall Length ........................................................................................................... 9ft. 9in (2.97m)
Sterilization Method .................................................................................................... EtO or Steam
Supplied Non-sterile
Reusable
RF Generator
Dimensions ......................................................................... 12.2” x 15” x 4.1” (31cmx38cmx10.5cm)
Weight ........................................................................................................................13.0 lb. (5.9kg)
Operating Temperature Range ...................................................................................... 10ºC to 40ºC
Operating Relative Humidity ............................................................... 15% to 80% (non-condensing)
Transport & Storage Temperature Range ............................................. -34ºC (-29ºF) to 65ºC (149ºF)
Transport & Storage Relative Humidity ............................................... 10% to 85% (non-condensing)
Shipped Non-sterile
Nominal Voltage ..................................... 115 Volts RMS (UL classified) or 230Volts RMS 50/60Hz
Operating Range ..................... 90-130 VAC @ 50/60Hz (UL classified) or 200–240 VAC @ 50 Hz
Inlet Fuse .............................................................................. 4A, 250V, T-lag, 5x20mm,UL/CSA or
Leakage Currents .......................................Within limits of Class CF equipment as per IEC 60601-1
Alarm Volume ........................ Adjustable between 45dB (minimum) and 65dB (maximum) at 1meter.
Intermittent Operation..............................
TECHNICAL SPECIFICATIONS
19
2.5A, 250V, T-lag, 5x20mm, VDE/CUL
Under maximum power setting and rated load conditions,
the Generator is suitable for 15 seconds on, 45 seconds
off duty cycle for 1 hour.
RF Output
Waveform.............................................460kHz±1%, square wave, Crest factor <1.3 @ 140 ohms
Maximum Power (CUT Mode)....................................................................225 Watts @ 140 ohms
Maximum Power (COAG Mode)...................................................................40 Watts @ 140 ohms

Other manuals for Stryker SERFAS