FUSE REPLACEMENT INSTRUCTIONS:
1. Unplug RF Generator from electrical outlet and unplug power cord from Generator.
2. Locate Fuse Box on rear panel of the Generator.
3. Unlatch fuse holder below the AC Inlet and remove. You may need to press the tab on the fuse holder with a
slender screwdriver to release the latch.
4. Replace the fuse with a new fuse of the same value rating as indicated on the back panel label.
WARNING: To avoid risk of fire, replace only with a fuse of the value specified on the fuse label on back of the RF
Generator.
5. Reinstall the fuse holder until the tab snaps in place.
HANDPIECE CABLE O-RINGS:
Silicone O-rings are inserted into the connectors on both ends of the Handpiece Cable to help prevent fluids from
reaching electrical contacts during use. The Cable O-rings may break or damage after repeated use. Inspect O-
rings periodically to ensure that the are still in place and are not damaged.
NOTE: Refer calibration and operating difficulties not detailed in this manual to your Stryker Endoscopy sales
representative, or Stryker Endoscopy Customer Service (in the United states: 1-800-624-4422 / 408-754-
2000). For service outside of the United States, contact your local Stryker Endoscopy Distributor (listed on
pages 24-25).