The RF Generator cannot be sterilized. If the Generator needs to be cleaned, wipe it down with a damp cloth or sponge.
Use only non-abrasive cleaning agents and DO NOT allow liquid to enter connectors on Generator.
FOOFOO
FOOFOO
FOO
TSWITTSWIT
TSWITTSWIT
TSWIT
CHCH
CHCH
CH
Wipe the outside of the SERFAS Footswitch with a mild detergent and water according to standard hospital practices.
Do not disassemble the SERFAS Footswitch. Do not sterilize the SERFAS Footswitch. Do not immerse the SERFAS
Footswitch in liquid. Use only non-abrasive cleaning agents and DO NOT allow liquid to enter connectors.
PROBESPROBES
PROBESPROBES
PROBES
Probes are shipped sterile and are intended to be used ONE TIME ONLY. Do not attempt to resterilize or reuse Probes.
Dispose of Probes after single use.
HANDPIECE CABLEHANDPIECE CABLE
HANDPIECE CABLEHANDPIECE CABLE
HANDPIECE CABLE
CLEANING:
The Handpiece Cable is shipped NON-STERILE and must be cleaned and sterilized before its first use and before each
subsequent use. Use the following cleaning and sterilization instructions.
1. Remove any gross matter (blood, mucous, tissue) by wiping with a soft cloth or gauze pad and by using a non-
abrasive, mild cleanser according to hospital standards.
2. The cable can be rinsed under running water to remove any cleaning residue, but should not be immersed in water
or any other liquid.
3. After cleaning, inspect handpiece cable for any signs of damage, such as nicks or cuts. DO NOT reuse a damaged
cable.
4. Inspect the inside of the connectors at both ends of the cable to ensure that the O-rings are still in place or for any
signs of damage.
CLEANING AND STERILIZATION
14